Surgical and Operating Room Equipment

   
   
Sterilization
 

Midmark Ritter M-9

 
Pulse-air Removal Technology
The M9's Steam-Flush Pressure-Pulse Air Removal System is so efficient that it can pass the air removal tests required on more complex vacuum steam autoclave systems, thus ensuring good steam penetration for those challenging loads.
 
Convenience at a Glance

The LCD display indicates cycle selected, cycle temperature and exposure time. When the cycle enters sterilization mode, the temperature, pressure, and the remaining cycle time is displayed. Remaining cycle time is also displayed in the dry mode. Note: Temperature & pressure are not displayed during dry mode.

No More Wet Instruments

Once the sterilization is complete, the door of the autoclave opens automatically and quietly to dissipate steam and dry your instruments. You'll have dry, sterile instruments every time.

 

Safety Features

If the door is not closed completely or the reservoir water level is low, the LCD display will provide the appropriate message.  In either case, sterilization will not continue until you've attended to the unit.

   
 

Ease of Draining & Filling

For ease of draining and filling, the M9's reservoir fill port and drain tube are in the front of the unit, not on top or in back.

   
 

Optional Printer Accessory

The M9 printer accessory provides a permanent record of cycle times,  temperatures and pressures during a cycle.

   

Self Program Controls

Programmable controls allow for creation of different cycle parameters for special sterilization needs. Programmable buttons allow you to change the following:  Exposure Time (3 to 90 minutes), Temperature (230 to 275º F), Dry Time (0 to 60 minutes), or Vent (slow or fast). The #1 and #2 buttons allow you to store the change if desired.

Pre-Programmed Controls

Four pre-programmed sterilization cycle controls: Unwrapped (132º C or 270º F for 3 minutes), Pouches (132º C or 270º F for 5 minutes), Packs (121º C or 250º F for 30 minutes), Hand pieces (unwrapped and pouched) (132º C or 270º F for 6 minutes).

LCD Display

Indicates cycle selected, temperature and exposure time. During the cycle, the display shows messages describing status of cycle. When the cycle enters sterilization mode, remaining time is displayed as well as temperature and pressure.

   
  Ultrasonic Units  (click here)
   
     
 

Should you have questions or concerns, please contact us

 574-268-9098 or sales@didage.com

 
 

The technical data provided is for general information and is subject to change without notice.